7-62
Performance
Check
Procedure
Connect
calibration
generator
(PG
506)
Std
Ampl
output
to
the
CH
1
and
the
EXT
Z-AXIS
Inputs
via
a
50
precision
BNC
coaxial
cable,
a
BNC
T-connector,
and
two
50
2
BNC
coaxial
cables.
Set
generator
to
Std
Ampl
output,
5
V,
Set
A
INTEN
to
maximum
intensity.
CHECK—waveform
display
Intensity
starts
decreasing
at
1.8
V
or
less
and
the
waveform
Is
completely
blanked
out
at
3.8
V.
Set
A
INTEN
to
midrange.
Disconnect
the
test
equipment
from
the
2247A.
PROBE
ADJUST
Output
Set:
CH
1
Vertical
MODE
10
mV
SEC/DIV
0.2
ms
Connect
a
10X
probe
to
the
CH
1
Input
connector
and
connect
the
probe
tlp
to
the
2247A
PROBE
ADJUST
output.
(When
using
Tektronix
coded
probes
the
readout
changes
to
.1V.)
CHECK—for
a
5-division
vertical
display
of
PROBE
ADJUST
square-
wave
signal
(square-wave
period
Is
typically
1
ms,
within
25%).
AUTO
SETUP
Functional
Check
Set:
CH
1
COUPLING
GND
CH
1
VOLTS/DIV
2mvV
A
SEC/DIV
20
ns
Press
the
AUTO
SETUP
button.
Check
that
the
Probe
Adjust
waveform
Is
stably
displayed
on
the
upper
half
of
the
crt.
2247A
Operators
c-3
1
ES
HS
CS