Performance verification
Harmonic Distortion Test Record (cont.)
Harmonic Distortion
(amplitude 1 V
p-p
)
Fundamental
= reference 2nd 3rd 4th 5th Limit
Sine 100 kHz
100 kHz 200 kHz 300 kHz 400 kHz 500 kHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-60 dBc
CH2 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-60 dBc
Sine 1 MHz
1MHz 2MHz 3MHz 4MHz 5MHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-50 dBc
CH2 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-50 dBc
Sine 5 MHz
5 MHz 10 MHz 15 MHz 20 MHz 25 MHz
CH1 H armonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-40 dBc
CH2 H armonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-40 dBc
Sine 25 MHz
25 MHz 50 MHz 75 MHz 100 MHz 125 MHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-40 dBc
CH2 Harmonic
Distortion (dBc)
AFG3000 Series Specifications and Performance Verification 37