Performance verication
Harmonic Distortion Test Record (cont.)
Harmonic Distortion
(amplitude 1 V
p-p
)
Fundamental
= reference 2nd 3rd 4th 5th Limit
reading -
reference
(dBc)
0 dBc
Nth-reference<
-37 dBc
CH2 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-37 dBc
Sine 100 MHz
100 MHz 200 MHz 300 MHz 400 MHz 500 MHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-37 dBc
CH2 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth-reference<
-37 dBc