EasyManua.ls Logo

Tektronix DPO3034 - Trigger Setup; Triggering Concepts

Tektronix DPO3034
302 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Trigger Setu p
Trigger Setup
This section contains concepts and procedures for setting up the oscilloscope to trigger o n your signal.
Triggering Concepts
Trigger Event
The trigger event establishes the time-reference point in the waveform record. All waveform record data is located in time with
respect t o that point. The instrument continuously acquires and retains enough sample points to ll the pretrigger portion of th e
waveform record. That is the part of the waveform that is displayed before , or to the left of, the triggering event on screen. When a
trigger event occurs, the instrument starts acquiring samples to build the posttrigg er portion of the w aveform record, that is, the part
displayed after or to the right of the trigger event. After a trigger is recognized, the instrument will not accept another trigger until
the acquisition is complete and the holdoff time has expired.
DPO3000 Series Oscilloscopes User Manua l 107

Table of Contents

Other manuals for Tektronix DPO3034