8. Record the peak-to-peak measurement as follows:
a. Push the front-panel Wave Inspector Measure button.
b. In the Measurements box, click Pk-Pk.
c. This will provide a mean V
p-p
of the signal. Call this value V
in-pp
.
d. Enter this value in the test record.
9. Set the Horizontal Scale to 5 ns/division.
10. Adjust the signal source to the maximum bandwidth frequency for the
bandwidth and model desired, as shown in the following worksheet.
11. Record the peak-to-peak measurement as follows:
a. With the Pk-Pk measurement still displayed from step 8, view the
measurement at the new frequency.
b. This will provide a mean V
p-p
of the signal. Call this value V
bw-pp
.
Performance verification (MSO/DPO5000/B series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 327