Application Exa
mples
5. Find glitches,
transients, or other random
anomalies that are present in the signal.
In this example, FastAcq revealed an
300 ns positi
ve glitch after only a few
seconds.
6. To trigger on the glitch that you identied
in step 5, s
elect Glitch Setu p....
7. Select the
appropriate polarity.
8. Click Lev
el, and then set the level based
on what you found in step 5.
9. Click Width, and then set the width
based on what you found in step 5.