Click the Trig After Time button under A Then B.
Click the B Trig Level control in the control window.
Set the front-panel trigger mode to Norm.
b. Confirm that the following statements are true:
The trigger-level readout for the B trigger system changes as you
turn the lower multipurpose knob.
As you rotate the lower multipurpose knob, the square-wave signal
can become triggered and untriggered. (Leave the signal triggered.)
c. Verify the delayed trigger counter:
Double click the Trig Delay control to pop up a numeric keypad for
that control.
Click the keypad to enter a trigger delay time of 1 second (click
1 and None) and then click Enter.
Verify that the trigger Ready indicator on the front panel flashes
about once every second as the waveform is updated on-screen.
7. Remove the test hookup: Disconnect the test hookup from the instrument.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
110 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series

Table of Contents

Other manuals for Tektronix DSA70000D Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000D Series and is the answer not in the manual?

Tektronix DSA70000D Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000D Series
CategoryTest Equipment
LanguageEnglish

Related product manuals