Repeat substeps 2.c through 2.f until all vertical scale settings are
checked for the channel under test. (See Table 7: DC voltage
measurement accuracy on page 184.)
g. Test all channels: Repeat substeps 2.a through 2.f for all four channels.
h. If testing <4 GHz models, set the DC Calibration generator to drive a
load impedance of 1 MΩ. Repeat substeps 2.a through 2.g with the
instruments termination set to 1 MΩ.
3. Disconnect the hookup:
a. Set the generator output to 0 V.
b. Disconnect the equipment from the generator output and the input
connector of the channel last tested.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 187