bandwidth test hookup, < 3.5 GHz models on page 228.) Set the output
of the generator to a reference frequency of 50 MHz or less.
Figure 20: Analog bandwidth test hookup, < 3.5 GHz models
2. Confirm the input channels are within limits for analog bandwidth: Do the
following substeps - test Ch 1 first, skipping substeps 2.aand 2.bsince Ch 1 is
already set up for testing from step 1.
a. Select an unchecked channel:
From the button bar, touch Measure and then Clear All to remove
the previous measurement.
Press the Vertical button of the channel just confirmed to remove the
channel from the display.
Press the front-panel Vertical button that corresponds to the channel
you are to confirm.
Touch Vertical, select Vertical Setup, and then touch Termination
50 Ω.
Move the leveled output of the sine wave generator to the channel
you selected.
b. Match the trigger source to the channel selected: Press the Trigger
Source button until the source that corresponds to the channel you are to
confirm is on.
Table 12: Analog bandwidth, < 3.5 GHz models
Vertical scale Reference
amplitude
Horizontal scale Test frequency -3 dB
DPO7254C DPO7104C DPO7054C
5 mV 30 mV 1 ns 2.0 GHz 1 GHz 500 MHz ≥21.2 mV
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
228 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series

Table of Contents

Other manuals for Tektronix DSA70000D Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000D Series and is the answer not in the manual?

Tektronix DSA70000D Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000D Series
CategoryTest Equipment
LanguageEnglish

Related product manuals