3. Record the reference level: Note the reading on the level meter.
4. Set the generator to the new frequency and reference level:
Change the sine wave generator to the desired new frequency.
Input the correction factor and/or the new frequency into the level meter.
Adjust the sine wave generator amplitude until the level meter again
reads the value noted in step 3. The signal amplitude is now correctly set
for the new frequency.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
282 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series