P6516 Digital Probe
Threshold Accuracy: ±(100 mV + 3% of threshold)
Maximum signal swing: 6.0 V
peak-to-peak
centered around the threshold voltage
Minimum signal swing: 500 mV
peak-to-peak
Maximum nondestructive input signal to probe: ±15 V
Input resistance: 20 K
Input capacitance: 3.0 pF typical
Temperature:
Operating: 0 °C to +50 °C (+32 °F to +122 °F)
Nonoperating: -55 °C t o +75 °C (-67 °F to +167 °F)
Altitude:
Operating: 4.5 km (15,000 ft) maximum
Nonoperating: 15 km (50,000 ft) ma ximum
Pollution Deg ree: 2, Indoo r use only