Installation
5. Push Autoset. The screen should now
display a square wave, approximately
2.5Vat1kHz.
NOTE. For best performance, it is
recommended that yo u set the Vertical scale
to 50 0 mV.
If the s ignal appears but is misshap en,
perform the procedures for compensating
the pro be. (See page 23, Compensating a
Passive Voltage Probe.)
If no signal app ears, rerun the procedure .
If th is does not remedy the situation, have
the instrument serviced by qualied service
personnel.