Remove the rear chassis assembly.
Procedure:
1. Disconnect the line input cable from the power supply
.
2. Use a T
-10 Torx screwdriver to remove the seven screws from the power supply.
3. T
o reinstall, reverse the steps. Tighten the T-10 Torx screws to 0.65 N·m when reinstalling.
Remove the main fan assembly
The following procedure describes the removal and replacement of the power supply bracket.
Prerequisite:
To prevent electrostatic damage to components whenever you work on the instrument, wear properly-grounded electrostatic prevention
wrist and foot straps, and work in a tested antistatic environment on an antistatic mat.
Remove the rear grill and case.
Remove the rear chassis assembly.
Procedure:
1. Use a T-10 Torx screwdriver to remove the seven screws from the main fan assembly bracket.
Maintenance
4 Series MSO (MSO44, MSO46) Service 26

Table of Contents

Other manuals for Tektronix MSO44

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO44 and is the answer not in the manual?

Tektronix MSO44 Specifications

General IconGeneral
Bandwidth200 MHz, 350 MHz, 500 MHz, 1 GHz
Analog Channels4
Operating Temperature0 °C to +50 °C
Sample Rate6.25 GS/s
Record Length62.5 Mpoints
Input Impedance1 MΩ ±1%, 50 Ω ±1%
Vertical Resolution12 bits
Waveform Capture RateUp to 500, 000 wfm/s
Maximum Input Voltage5 VRMS (50 Ω)
Probe InterfaceTekVPI
Digital Channels16 (optional)

Related product manuals