How to clear or sanitize a working instrument
How to clear or sanitize a working instrument
Run all the following procedures, in order, to clear or sanitize the instrument. You typically clear or sanitize an instrument
when you want to erase files to clear space or turn the instrument over to another person or department.
Clear the Network Configuration password
1.
Connect the instrument to a n etwork to which you have access.
2.
Enter the instrument’s IP address into a web browser on a PC that has network access to the instrument.
3.
Click the
Security for Network Config
link on the left side of the screen.
4.
Click
Submit
:
If a password was set for this function, you are requested to enter the password. If the password is accepted, the
password is set to blank (the default setting of the access password fields).
If a password was not set for this function, the screen displays the message that the password was successfully
changed (to a blank password).
Clear the network DNS Hostname and description
1.
Connect the instrument to a n etwork to which you have access.
2.
Enter the instrument’s IP address into a web browser on a PC that has network access to the instrument.
3.
Click the
Network Configuration
link on the left side of the screen.
4.
Delete all text in both of the
Host Settings
fields.
5.
Click the Host Settings
Submit
button. A message appears stating that the field is empty, and will be configured to
the original factory default value.
6.
Click
OK
. The message closes and the fields are restored to their original factory settings.
10 MSO58LP with Option 5-SEC Declassification and Security Instructions