NOTE. The MagniVu record must always contain the trigger. If the trigger point is moved prior to the display region using
Horizontal Delay, then the MagniVu record will not be available. In most of these cases, you can find the digital record by looking
for the digital signal in the upper overview and panning accordingly.
NOTE. You should turn on MagniVu when light gray shading is displayed to indicate the uncertainty of the edge position. If the
shading is not displayed, you do not need to use MagniVu.
MagniVu setup
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series 165