TIP. Actions for violations
The analyzer can take up to five actions when a match occurs. In Run mode, the analyzer can sound a
beep, stop acquisitions, save acquisition data, save trace data, and save a picture of the display. In
Replay mode, it can sound a beep and stop analyzing.
Files can be saved in the native format of SignalVu-PC (.tiq), or in comma separated variable (.CSV) or
Matlab level 5 (.MAT) formats.
The folder for saving files can be selected from Save as….