TAS 200 Series Adjustment Procedures
26
Bench Test Instruments and Handheld Oscilloscopes Basic Service
Following the warm-up period, preset the oscilloscope to the settings listed
below.
Table 20: Oscilloscope initial settings
Control Setting
INTENSITY Visible displ ay
CURSOR ON/OFF Off
VERTICAL
CH 1 (CH 2) POSITION Midrange
MODE CH1 (CH2)
CH 1 (CH 2) VOLTS/DIV 10 mV
CH 1 (CH 2) VOLTS/DIV VARIABLE CAL
CH 1 (CH 2) GND In
HORIZONTAL
POSITION Midrange
X10 MAG Out
X-Y Out
SWP UNCAL Out
TRIGGER
LEVEL Midrange
MODE AUTO
COUPLING AC
SOURCE CH 1 (CH 2)
SLOPE Rising (push
switch out )
Initial Settings