EasyManuals Logo

Tektronix TAS 250 Technical Reference

Tektronix TAS 250
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #378 background imageLoading...
Page #378 background image
THM500 Series Performance Verification
22
Handheld and Benchtop Instruments Basic Service
THM5xx
CH 2COM
CH 1
50 termination
50 cable Patch cord
Leveled Sine
Wave
generator
BNC to dual banana cable
Figure 10: Crosstalk Between Scope Channels Test Setup
3. On the THM500 series instrument, press the POSITION/SCALE button
until the readout shows SCALE in the lower-right corner. Use the
Y
and
B
buttons to set the CH 1 vertical scale to 500 mV/div and the
A
and
"
buttons to set the time base to 200 ms/div.
4. Use the THM500 series instrument menus to set the sampling mode to
SPIKE DETECT; then remove the menus from the display.
5. Adjust the leveled sine wave generator to display five divisions peak-to-peak
at 5 MHz.
NOTE. The 5 MHz waveform is aliased and appears untriggered.
6. Press the THM500 series instrument WAVEFORM ON/OFF button to turn
the CH 1 waveform display off.
7. Press the THM500 series instrument SELECT button to select CH 2. Press
the WAVEFORM ON/OFF button to turn on channel 2. Press the POSI-
TION/SCALE button to show SCALE; then use the
Y
and
B
buttons to set
the vertical scale to 50 mV/div.
8. Verify that the peak-to-peak signal displayed on the THM500 series
instrument channel 2 is less than 0.5
divisions (25 mV
p-p
).
9. Disconnect the test setup.

Table of Contents

Other manuals for Tektronix TAS 250

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TAS 250 and is the answer not in the manual?

Tektronix TAS 250 Specifications

General IconGeneral
BrandTektronix
ModelTAS 250
CategoryTest Equipment
LanguageEnglish

Related product manuals