EasyManua.ls Logo

Tektronix TAS 250 - Page 24

Tektronix TAS 250
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TAS 200 Series Performance Verification
20
Bench Test Instruments and Handheld Oscilloscopes Basic Service
6. Position the O cursor 3 divisions below the display horizontal center with the
CURSOR POSITION switch.
7. Verify that the V
1(2)
readout at the top-left corner of the display, reads
between 5.78 and 6.22 V.
8. Set the VERTICAL MODE switch to CH2 and repeat steps 3 through 7 for
CH 2.

Other manuals for Tektronix TAS 250

Related product manuals