EasyManua.ls Logo

Tektronix TDS 520B - Figure 4-21: Measurement of Main Trigger out Limits

Tektronix TDS 520B
311 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests
4–60
TDS 500B, 600B and TDS 700A Service Manual
Check output
Figure 4–21: Measurement of Main Trigger Out Limits
d. Check Delayed Trigger output against limits: See Figure 4–21.
H Move the precision 50 W cable from the rear-panel Main Trigger
Output BNC to the rear-panel Delayed Trigger Output BNC.
H CHECK that the Ch2 High readout is 1.0 volt and that the Ch2
Low readout 0.25 volts.
H Enter high and low voltages on test record.
H Press the side-menu button to select the 1 MW setting.
H Press CLEAR MENU.
H CHECK that the Ch2 High readout is 2.5 volts and that the Ch2
Low readout is 0.7 volts.
H Enter high and low voltages on test record.
3. Confirm CH 3 (Ax1 on TDS 520B, 620B, 680B, and TDS 724A) output is
within limits for gain:
a. Measure gain:
H Move the precision 50 W cable from the rear-panel DELAYED
TRIGGER OUTPUT BNC to the rear-panel SIGNAL OUT BNC.

Table of Contents

Other manuals for Tektronix TDS 520B

Related product manuals