Removal and Installation Procedures
6–20
TDS 500B, 600B and TDS 700A Service Manual
b. Take care not to bind or snag the cabinet on internal cabling; redress
cables an necessary.
c. When sliding the cabinet, be sure that the front edge of the cabinet aligns
with the groove containing the EMI shield on the front casting.
d. Be sure that the ridge around the rear chassis slides into the groove
containing a second EMI cable on the rear of the cabinet.
e. When reinstalling the four screws at the rear panel, tighten them to
16 foot-lbs torque.
f. See the procedure Line Fuse and Line Cord to reinstall the line cord,
which completes the oscilloscope reassembly.