Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–91
H Press HORIZONTAL MENU; then press the main-menu button
Trigger Position. Set the Trigger Position to 20% using the general
purpose knob or keypad.
NOTE. Press SHIFT to change the horizontal position speed. The position moves
faster when SHIFT is lighted.
H Adjust the HORIZONTAL POSITION to move the 20% trigger
point onto the display (see Figure 1–44).
Figure 1–44: Optical impulse of Ch1 input from OA5022 Optical Attenuator
(OIG501 / OIG502 fed into optical attenuator in Step 1)
j. Select linear interpolation:
H If SHIFT is lighted, press SHIFT.
H Press DISPLAY; then press the main-menu button Filter.
H Press the side-menu button Linear Interpolation.
2. Select the desired Mask standard and turn it on:
H Press MEASURE; then select Masks from the Measure pop-up menu.