Performance Tests
1–60
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
Check output
Figure 1–21: Measurement of main trigger out limits
d. Check Delayed Trigger output against limits: See Figure 1–21.
H Move the precision 50 W cable from the rear-panel Main Trigger
Output BNC to the rear-panel Delayed Trigger Output BNC.
H CHECK that the Ch2 High readout is 1.0 volt and that the Ch2
Low readout 0.25 volts.
H Enter high and low voltages on test record.
H Press the side-menu button to select the 1 MW setting.
H Press CLEAR MENU.
H CHECK that the Ch2 High readout is 2.5 volts and that the Ch2
Low readout is 0.7 volts.
H Enter high and low voltages on test record.
3. Confirm CH 3 (Ax1 on some TDS models) output is within limits for gain:
a. Measure gain:
H Move the precision 50 W cable from the rear-panel DELAYED
TRIGGER OUTPUT BNC to the rear-panel SIGNAL OUT BNC.
H Push TRIGGER MENU.