Maintenance
TDS1000 and TDS2000 Series Digital Storage Oscilloscopes Service Manual
6-- 45
Clearing the Error Log. To clear the error log after fixing a problem, follow these
steps:
1. Enable the Service menus as described on page 5--3.
2. Push the UTILITY button and push the Service option button to display the
Service menu.
3. Push the Service Diag. option button to display the Diagnostics menu.
4. Push the Clear Error Log option button to clear the error log of past
failures. This also sets the Power Cycle count to zero.

Table of Contents

Other manuals for Tektronix TDS1002

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1002 and is the answer not in the manual?

Tektronix TDS1002 Specifications

General IconGeneral
Bandwidth60 MHz
Channels2
Sample Rate1 GS/s
Record Length2.5k points
Vertical Sensitivity2 mV/div to 5 V/div
Vertical Resolution8 bits
Display Resolution320 x 240 pixels
Display5.7 inch monochrome LCD
Trigger ModesVideo, Edge
ConnectivityUSB 2.0
Weight2.0 kg
Input Impedance1 MΩ
Power Supply100 to 240 VAC, 45 to 440 Hz