Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−49
CALibrate:INTERNAL (No Query Form)
Performs an internal self-calibration but does not return any status.
This is equivalent to selecting the Do Self Cal option in the Utility
menu.
NOTE. The self-calibration can take several minutes to complete.
During this time, the oscilloscope does not execute any commands.
Disconnect all signals from the oscilloscope before performing an
internal self-calibration.
CALibrate:INTERNAL
CALibrate:INTERNAL
performs an internal self-calibration.
*CAL?
CALibrate:STATUS? (Query Only)
Returns the status of the last calibration operation performed (either
self- or factory-calibration) since power up.
Calibration and Diagnostic
CALibrate:STATUS?
PASS indicates that the oscilloscope completed the last calibration
operation without detecting any errors.