Specifications
1-4
TDS1000B and TDS2000B Series Oscilloscope Service Manual
Table 1- 1: Oscilloscope specifications (Cont.)
Vertical w
n DC Gain Accuracy
±3% for Sample or Average acquisition mode, 5 V/div to 10 mV/div
D
C
G
a
i
n
A
c
c
u
r
a
c
y
±4% for Sample or Average acquisition mode, 5 mV/div and 2 mV/div
Volts Measurement
Repeatability, Average
Acquisition Mode
Delta volts between any two averages of
16 waveforms acquired under same setup
and ambient conditions
±(3% × reading + 0.05 div)
DC Measurement Measurement Type Accuracy
Accuracy, Average
Acquisition Mode
Average of 16 waveforms with vertical
position at zero
±(3% × reading + 0.1 div + 1 mV) when 10 mV/div or
greater is selected
Average of 16 waveforms with vertical scale
with Vertical Scale 2 mV/div to 200 mV/div and
-1.8 V < Vertical Position < 1.8V
±[3% × (reading + vertical position) + 1% of vertical
position + 0.2 div + 7 mV]
Average of 16 waveforms with vertical
position with Vertical Scale > 200 mV/div and
-45 V < Vertical Position < 45 V
±[3% × (reading + vertical position) + 1% of vertical
position + 0.2 div +175 mV]
Horizontal
Sample Rate Range TDS1001B, TDS1002B, TDS2004B,
TDS1012B, TDS2002B, TDS2012B, TDS2014B
TDS2022B, TDS2024B
5S/sto1GS/s 5S/sto2GS/s
Waveform Interpolation (sin x)/x
Record Length 2500 samples for each channel
SEC/DIV Range TDS1001B, TDS1002B, TDS2004B,
TDS1012B, TDS2002B, TDS2012B, TDS2014B
TDS2022B, TDS2024B
5 ns/div to 50 s/div, in a 1, 2.5, 5 sequence 2.5 ns/div to 50 s/div, in a 1, 2.5, 5 sequence
n Sample Rate and
Delay Time Accuracy
±50 parts per million (ppm) over any 1 ms time interval
Delta Time Measurement Conditions Accuracy
Accuracy (Full Band-
w
i
d
t
h
)
Single-shot, Sample m ode ±(1 sample interval + 100 ppm × reading + 0.6 ns)
w
i
d
t
h
)
> 16 averages ±(1 sample interval + 100 ppm × reading + 0.4 ns)
Sample interval = s/div ÷ 250
Position Range TDS1001B, TDS1002B, TDS2004B, TDS1012B, TDS2002B, TDS2012B, TDS2014, TDS2022B, TDS2024B
g
5ns/divto10ns/div (--4 div × s/div) to 20 ms
25 ns/div to 100 s/div (--4 div × s/div) to 50 ms
250 s/divto10s/div (--4 div × s/div) to 50 s
25 s/div to 50 s/div (--4 div × s/div) to 250 s
TDS2022B, TDS2024B
2.5 ns/div (--4 div × s/div) to 20 ms
w Specifications are with the Probe " Voltage " Attenuation option set to 1X.

Table of Contents

Other manuals for Tektronix TDS2022B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2022B and is the answer not in the manual?

Tektronix TDS2022B Specifications

General IconGeneral
BrandTektronix
ModelTDS2022B
CategoryTest Equipment
LanguageEnglish

Related product manuals