Specifications
TDS1000B and TDS2000B Series Oscilloscope Service Manual
1-3
Table 1- 1: Oscilloscope specifications (Cont.)
Vertical w
Digitizers 8-bit resolution (except when set to 2 mV/div), each channel sampled simultaneously
VOLTS/DIV Range 2 mV/div to 5 V/div at input BNC
Position Range, typical 2 mV/div to 200 mV/div, ±2V
> 200 mV/div to 5 V/div, ±50 V
n Vertical Position
Measurement Type Volts/Div Setting Position Accuracy
Accuracy
Accuracy of the nominal voltage level
represented by the code at the center of
the dynamic range of the A/D converter
2 mV/div to 200 mV/div ±(1% X selected value + 0.1 div + 5 mV)
within the range ±1.8 V
W
> 200 mV/div to 5 V/div ±(1% X selected value + 0.1 div + 125
mV) within the range ±45 V
W
n Analog Bandwidth
in Sample and Average
TDS1001B TDS1002B, TDS2002B,
TDS2004B
TDS1012B, TDS2012B,
TDS2014B
TDS2022B, TDS2024B
modes at BNC or with
P2220 probe set to 10X,
DC Coupled
40 MHz*{
60 MHz*{ 100 MHz*{ 200 MHz*{
0 °Cto+35°C(32°Fto95°F)
160 MHz*{
0 °Cto+50°C(32°F to 122 °F)
20 MHz* (when vertical scale is set to < 5 mV)
Analog Bandwidth in
Peak Detect mode
TDS1001B TDS1002B, TDS2002B,
TDS2004B
TDS1012B, TDS2012B, TDS2014B, TDS2022B, TDS2024B
(50 s/div to 5 s/div
}
),
30 MHz*{ 50 MHz*{ 75 MHz*{
20 MHz* (when vertical scale is set to < 5 mV)
Selectable Anal og Band-
width Limit, typical
20 MHz*
Lower Frequency
Limit, AC Coupled
≤ 10 Hz at BNC
≤ 1 Hz when using a 10X passive probe
Rise Time at BNC,
typical
TDS1001B TDS1002B, TDS2002B,
TDS2004B
TDS1012B, TDS2012B,
TDS2014B
TDS2022B, TDS2024B
< 8.4ns <5.8ns <3.5ns <2.1ns
Peak Detect Response
}
Captures 50% or greater amplitude of pulses ≥12 ns wi de typical (50 s/div to 5 s/div) in the center 8
vertical divisions
For TDS1001B, the pulse must be at least 13 ns wide
w Specifications are with the Probe " Voltage " Attenuation option set to 1X.
W
The “selected value” is the offset value indicated by the oscilloscope in the hint line (not the measurement).
* Bandwidth reduced to 6 MHz with a 1X probe and the Bandwidth Limit is turned off.
{ When vertical scale is set to ≥ 5mV.
}
The oscilloscope reverts to Sample mode when the SEC/DIV (horizontal scale) is set from 2.5 s/div to 5 ns/div on 1 GS/s
models, or from 2.5 s/div to 2.5 ns/div on 2 GS/s models. The Sample mode can still capture 10 ns glitches.