HVG-2020B
INSTRUCTION MANUAL
I S O 9 0 0 1
C E R T I F I E D
www.teledyne
-
hi.com
Use and Disclosure of Data
Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations.
Export, reexport or diversion contrary to U.S. law is prohibited.
Question and Answer IconNeed help?

Do you have a question about the Teledyne HVG-2020B and is the answer not in the manual?

Overview

The Teledyne Hastings HVG-2020B is a vacuum gauge designed for precise pressure measurement across a wide range, from 1000 Torr down to 10⁻⁴ Torr. It combines a Pirani sensor and a piezo-resistive sensor to provide accurate and stable readings, with the piezo-resistive sensor offering gas species independence for direct pressure measurement. The device is available with an optional integrated touch panel color LCD display, enabling easy configuration, stand-alone operation, and local control.

Function Description:

The HVG-2020B measures vacuum pressure using a dual sensor system. The piezo-resistive sensor provides accurate absolute pressure measurements from 1000 Torr to 1 Torr, independent of gas composition. The Pirani sensor extends the measurement range down to 10⁻⁴ Torr. The instrument features thermal compensation and linearization for enhanced accuracy. It offers various output options, including analog voltage or current signals (logarithmic or linear scale), and digital communications via RS232/RS485 and USB. Two open-drain TTL alarm signals are provided to indicate when pressure exceeds or falls below predefined values. The device is designed for robust performance in industrial environments, with a 300 series stainless steel vacuum envelope, welded construction, and glass-to-metal seals, eliminating the issues associated with elastomer O-rings.

Important Technical Specifications:

  • Range: 1x10⁻⁴ Torr to 1000 Torr
  • Accuracy:
    • ± (0.1% of reading + 0.3 Torr) from 1000 to 20 Torr
    • ± (5% of reading + 0.00025 Torr) from 20 to 0.001 Torr (Nitrogen)
    • ± (10% of reading + 0.00025 Torr) from 0.001 to 0.0001 Torr (Nitrogen) at ambient operating temperature without temperature/humidity effects after 2 hours warm up followed by zero adjustment.
  • Warm up time: Within rated accuracy at atmosphere: 30 minutes (typical). For zero adjustment: 2 hours (typical).
  • Installation orientation: Any
  • Internal volume: 3.5 cc (1/4" stub tube, other connections will be larger)
  • Maximum Pressure: 1000 Torr
  • Maximum overpressure: 2000 Torr / 25 psig (no change in performance)
  • Burst pressure: 3000 Torr / 45 psig
  • Operating Temperature: -20°C to 70°C
  • Humidity: 30 to 90 %RH, non-condensing
  • Thermal Zero Shift: < 3.5 mTorr (over -20 to 70°C range)
  • Thermal Span Shift:
    • < 0.6% of reading above 10 Torr (over -20 to 70°C range)
    • < 5% of reading below 10 Torr (over -20 to 70°C range)
  • Leak Integrity: < 1x10⁻⁹ std. cc/s
  • Analog Output:
    • 0-1 VDC, 0-5 VDC, 0-10 VDC (short-circuit protected, 1 kΩ min load)
    • 0-20 mA, 4-20 mA (short-circuit protected, 1.2 kΩ max load)
  • Power Requirements:
    • With display: 2.1 watts max. / 12 to 36 VDC (1.5 W typ. @ 24 VDC)
    • No display: 1.8 watts max. / 12 to 36 VDC (1.3 W typ. @ 24 VDC)
    • Class 2 power 150 VA max fused
  • Wetted Materials: 304 and 316 stainless steel, glass, nickel, gold plated tungsten
  • Weight (approx.): 0.6 lb. (0.27 kg) without fitting
  • Set points: Open-drain configuration (5 V / 5 mA max input)
  • Connectors:
    • 2mm bayonet-style power jack (power)
    • 9-pin "D" connector (power, analog signals, TTL serial)
    • 3.5 mm TRRS jack (RS232/485 serial communications)
    • USB micro-B (virtual COM port for serial communications)

Usage Features:

  • Color Display with Touch Panel: Provides an intuitive interface for configuration and monitoring. Users can view pressure, temperature, and other parameters, and adjust settings directly from the display.
  • Multiple Display Views: The VIEW icon allows users to cycle through five possible main screens (views), including Pressure View, Pressure and Temperature View, Setpoint View, Bar Graph View, and History Graph View.
  • MENU Button: Accesses a multipage ring of menu items for advanced settings, including Screen Orientation, Set Zero, Device Info, Analog Output, Gas Select, Serial Port, and Restore Factory Defaults.
  • Gas Correction: Internal corrections are available for the Pirani sensor when used with gases other than nitrogen, improving accuracy for specific applications.
  • Digital Communications: Equipped with RS232/RS485 and USB for seamless integration into control systems. The USB connector functions as a virtual COM port, eliminating the need for external converters.
  • Analog Outputs: Provides flexible analog output options (voltage or current) for easy integration with existing analog control systems. Current loop outputs (0-20 mA or 4-20 mA) minimize noise pickup and signal loss over long cable runs.
  • TTL Alarms: Two configurable alarm signals (open-drain) can be set to trigger when pressure crosses user-defined thresholds, providing critical process control feedback.
  • LED Indicators: Two LEDs provide quick visual feedback on instrument status and system pressure.
  • Quick Start Guide: Includes clear instructions for unpacking, installing, connecting, and performing initial setup, including zero adjustment.

Maintenance Features:

  • Calibration Procedure: Detailed steps for calibrating the HVG-2020B using actual gas, including adjustments for nitrogen and other specific gases.
  • Troubleshooting Guide: Provides symptoms and actions for common issues, such as communication problems, pressure gauge errors, and LED indications.
  • Authorized Maintenance: Emphasizes that maintenance should be performed by qualified personnel to ensure proper function and prevent damage.
  • Warranty Policy: Information on the company's warranty policy and return procedures is provided, along with contact details for repair and support.
  • Robust Construction: The use of 300 series stainless steel, welded construction, and glass-to-metal seals minimizes maintenance needs by preventing permeation, degradation, and outgassing issues common with elastomer O-rings.
  • Factory Defaults: Option to restore factory default settings, simplifying troubleshooting and re-configuration.

Teledyne HVG-2020B Specifications

General IconGeneral
BrandTeledyne
ModelHVG-2020B
CategoryMeasuring Instruments
LanguageEnglish