TM-197
5.1 Zeroing
Move the probe away from the magnetic field and
press , ZERO will appear on the LCD. After
zeroing completed, the ZERO disappears.
If more than 3G or 0.3mT, will
appear to prevent from erroneous
zeroing the value in measuring. (from
If more than 3G or 0.3mT, will
appear to prevent from erroneous
zeroing the value in measuring. (from
The zeroing function is unavailable for
5.2 Peak Measurement
Move the probe away from the magnetic field, then
press and hold for ≥2 seconds to start Peak
Measurement. The upper digit row presents the
instant measuring; the lower digit row presents the
peak value x.x
Again, press and hold to quit this function.
Zeroing will run automatically in prior to
Peak Measurement. If more than 3G or
0.3mT, will appear to prevent
from erroneous zeroing the value in