Tenmars TM-197 - Manual;Automatic Range Modes; AC;DC Switching magnetic field; Setup Procedure

Tenmars TM-197
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TM-197
8
5.12Manual range mode/Automatic range
mode
Press to operate in the manual range mode
with the LCD showing . This can change into two
different range modes.
Press and hold for 2 seconds to return to
automatic range mode with LCD showing .
5.13AC/DC Switching magnetic field
Press to switch the measurement between AC
magnetic field and DC magnetic field.
5.14SetupStep 1~ Step 8
Press to enter the setup procedure from step
1 to step 8, while flashes on the LCD.
Step 1. Setup unit G or mT.
1. Press or to select the unit G or mT.
2. Again, press to go to Step 2.
Step 2. Setup unit or
1. Press or to select the unit ºC or ºF.
2. Again, press to go to Step 3.
Step 3. Set up the upper limit (Max.) for alarm
GO/NO-GO
1. The default value is 20000G (2000.0mT). Press
or to modify and setup the default
value.
2. Again, press to go to Step 4.