Calibrating:
The configuration menu is open, config. is displayed.
1 Probe Calibr. .
2 Press / to select calibration point P1 or P2 and confirm by
pressing twice.
3 Put the humidity probe into the reference medium and wait for the
equalisation period to elapse.
- The current humidity reading and the calibration point (nominal value) are
displayed.
4 Start the calibration menu with .
5 Save calibration with or cancel calibration with .
P internal (435-3/-4 only)
The internal pressure sensor can be switched on/off.
Switching
the
internal
pressure
sensor
on/off:
The configuration menu is open, config. is displayed.
1 Probe P intern .
2 Press / On or Off and confirm with .
Te-Type
The probe characteristic curves stored in the instrument can be set for the
probe type used.
Setting
probe
type:
The configuration menu is open, Config. is displayed.
1 Probe Te-Type .
2 Select the desired probe type with / and confirm with .
Adjustment
The function is only available when an absolute pressure probe is attached.
The display value for the measurement of abolute pressure.
Carrying
out
adjustment:
The configuration menu is open, config. is displayed.
1 Probe Adjustm. .
2 With / , set the adjusted value and confirm with .
OK
OKOK
OK
OKOK
OK
OKOK
ESCOK
OK
OK
OKOK
6. Setting the instrument 17