K-factor
The querying of the K-factor (correction factor for areas) when entering the
parameter „Area“.can be switched on/off.
See chapter PARAMETER /AREA, p. 24.
Switching
K-ffactor
on/off:
The configuration menu is open, config. is displayed.
1 Device K-factor .
2 With / , select On or Off and confirm with .
Num holes
The querying of the number of holes (number of measurement points) in the
calculation of a multi-point mean value can be switched on/off. The number of
holes is required to allocate the readings to the number of measurement points
for later evaluation via the PC software.
See Chapter MEAN, p. 22 and MEASURING, p. 26.
Switching
number
of
holes
on/off:
The configuration menu is open, config. is displayed.
1 Device Num holes .
2 With / , select On or Off and confirm with .
6.1.4 Probe
RadioC
Radio probes may only be used in countries in which they have been Type
Approved (see application information of the radio probe).
A radio module (accessory part) is required for the use of radio probes. The
instrument can establish a connection with a maximum of three radio probes.
Each radio probe has a probe ID (RF ID). This consists of the last 3 digits of the
serial no. and the position of the slide switch (H or L) in the radio probe.
Setting
up
a
radio
probe:
A radio module (accessory part) is inserted in the instrument.
See chapter COMMISSIONING, p. 9.
The configuration menu is open, config. is displayed.
The radio probe is switched on and the transfer rate is set to 2 readings
per second (see the advice on using the radio probe).
OK
OKOK
OK
OKOK