6. Setting the instrument 19
Menu overview testo 435-1/-3
Profile menu items Function
Standard P = 0 (nur 435-3) Zero internal pressure sensor
Area Set form, cross-section area, K-factor
Calc. De/activate volume flow, differential temperature, dewpoint
temperature, psychrometric temperature calculation;
435-3 additionally; De/activate flow calculation
Parameter Set reference pressure, abs. altitude;
435-3 additionally; Set reference temperature/humidity.
cyc. Print De/activate cycle printing
Ductm. P = 0 (nur 435-3) Zero internal pressure sensor
Velocity (nur 435-3) De/activate flow calculation
Vol. De/activate volume flow calculation
Parameter (nur 435-3) Set reference pressure, abs. altitude, P-factor, Set reference
temperature/humidity.
Pres. (nur 435-1) Set reference pressure
cyc. Print De/activate cycle printing
Menu overview testo 435-2/-4
Profile menu items Function
Standard Memory Info, Activate/set measurement locality, print report, delete store
Meas. Prog Set/ de/activate measurement program
Mean Time/point mean calculation
Calc. De/activate volume flow, differential temperature, dewpoint tempera-
ture, psychrometric temperature, Enthalpy calculation; Set heat
transfer coefficient alpha
435-3 additionally; De/activate flow calculation
P = 0 (nur 435-4) Zero internal pressure sensor
Parameter Set reference pressure, abs. altitude, Set area form/cross-section;
435-4 additionally: Set reference temperature/humidity.
Ductm. P = 0 (nur 435-4) Zero internal pressure sensor
Memory Info, Activate/set measurement locality, print report, delete store
Velocity (nur 435-4) De/activate flow calculation
Vol. De/activate volume flow calculation
Parameter Set reference pressure, abs. altitude
435-4 additionally: Set reference temperature/humidity.
IAQ Mean Time mean calculation
Pres. Set reference pressure
Memory Info, Activate/set measurement locality, print report, delete store
abs alt. Set abs. altitude
P = 0 (nur 435-4) Zero internal pressure sensor