EasyManua.ls Logo

TESTO 880 - Questions and answers

TESTO 880
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Tips and assistance
38
8 Tips and assistance
8.1. Questions and answers
Question Possible causes/solution
Message Backup battery
empty or not present is
shown.
1. Confirm the message with
[OK].
2. Change buffer battery, see
Change buffer battery page 36.
Error No memory card
inserted! is shown.
Memory card defective or not
inserted.
1. Confirm the message with
[OK].
2. Check or insert memory card.
Error Memory card full! is
shown.
Insufficient memory present.
1. Confirm the message with
[OK].
2. Insert new memory card.
Error No lens/wrong lens!
is shown.
1. Switch off thermal imager.
2. Check lens
> Change lens if necessary.
Message Charging not
possible! is shown.
Ambient temperature is outside
the permissible range for charging
the battery.
1. Confirm the message with
[OK].
2. Keep to the permissible
ambient temperature.
Error Permissible
instrument temperature
exceeded! is shown.
1. Switch off thermal imager.
2. Let instrument cool off and
keep to the permissible
ambient temperature.
--- or +++ is shown instead
of the reading.
Reading is outside of the
measuring range.
> Change measuring range.

Other manuals for TESTO 880