EasyManuals Logo

TESTO Gas detector User Manual

TESTO Gas detector
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
J. Questions and Answers
This Chapter provides the answers to frequently asked questions
Question Possible causes Remedy
is shown in display Remaining life of rechargeable battery is Recharge battery (Refer to
less than15min Recharging battery, P. 9).
Detector switches itself Rechargeable battery empty Recharge battery (Refer to
off or detector cannot be Recharging battery, P. 9).
switched on.
E32 is shown in display Calibration data incorrect Send detector for repair to Customer
Service.
E64 is shown in display Sensor not connected Send detector for repair to Customer
Service.
E96 is shown in display Calibration data incorrect and sensor not Send detector for repair to Customer
connected Service.
E128 is shown in display Parameter memory defective Send detector for repair to Customer
Service.
E160 is shown in display Parameter memory defective and Send detector for repair to Customer
calibration data incorrect Service.
E192 is shown in display Parameter memory defective and sensor Send detector for repair to Customer
not connected Service.
E224 is shown in display Parameter memory defective and calibration data Send detector for repair to Customer
incorrect and sensor not connected Service.
If we were unable to answer your question. please contact your distributor or Testo
Customer Service. You will find contact details in the
Representatives and Worldwide
Guarantee
document or in Internet at
www.testo.com
.
J. Questions and Answers
15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TESTO Gas detector and is the answer not in the manual?

TESTO Gas detector Specifications

General IconGeneral
BrandTESTO
ModelGas detector
CategoryOther
LanguageEnglish

Related product manuals