ES Series Lab Freezer 8
Operation
Controller Overview and Operation
- Sets the showed probe on display between Air and Load probe
- Shows the Maximum recorded temperature related to the showed
running probe.
- Scrolling UP options.
- Shows the Minimum recorded temperature related to the showed
running probe.
- Scrolling DOWN options.
- Press for 5 seconds to access Menu functions.
- Pressed with ENTER for 5 seconds, turns ON and OFF the controller.
- Pressed with MUTE alarm for 5 seconds, parameters class “C”
(Configuration) are accessed.
- Pressed for 5 seconds performs a reset of both the Air and Load
maximum and minimum temperatures.
- Press to turn on the light. (if applicable i.e. light fitted)
- It will stay on for 30 seconds and will go off automatically after this
time.
- Press to Mute the alarm buzzer
- Pressed for 5 seconds. Parameters class “F” (frequent) are accessed.
- Pressed with SERVICE for 5 seconds, parameters class “C”
(configuration) are accessed.
- Pressed with ENTER for 5 seconds, locks and unlocks the controller.
- Confirms the selected value/function.
- Pressed with SERVICE for 5 seconds, turns ON and OFF the
controller
- Shows the second compartment temperature in a Fridge/Freezer.
Table 3. Controller Settings