7270105 Rev. 1
Preventive Maintenance Kit 2270105
3110 Series Water Jacket CO
2 Incubator
(Models 3110, 3111, 3120, and 3121)
Included in this kit (1 each unless specified):
Tools required:
Common hand tools
Caution These procedures should only be performed by qualified
personnel.
Caution Turn unit off and disconnect it from power before beginning any
maintenance or service procedures.
Note There are two blower scroll configurations available in some
instances. Follow the instructions pertaining to your configuration.
Application
Model
Instruction Sheet
Lithium Grease 1.5 oz. #6 Solid Silicone Stopper
#6 Silicone Stopper, 1-1/4 Silicone Hose, 1/4 D x 1/8 W x 1.1 L
Blower Wheel, 3.5 x 1.5 CCW Silicone Hose, 1/4 D x 1/16 W x 1 L
Blower Wheel, 3.5 x 1.5 CCW “D” Teflon Washer, 0.31 ID x 0.625
Blower Plate Gasket HEPA Filter 5.5 x 1.5
CO2 Sensor Gasket Disposable Filter (3)
DCS IR Sensor Plate Gasket 3 x 4 1/4 Hose x 1/4 Hose Connector
Silicone Grommet, 3/8 x 1/4 (2) Corrosion Inhibitor, 800mL
Blind Silicone Grommet, 15/16 (2)