OPERATION: (Con’t)
TEMPERATURE CONTROLLER:
HEAT INDICATOR CHAMBER OR SETPOINT TEMPERATURE
65.0
1. CONTROLLER SELF-TEST: When the incubator is powered up the
controller will display 8888 along with
the three decimal points and the heat ON
indicator lamp. The display will then blank
out for 2 seconds before showing the
chamber temperature.
2. HEAT ON INDICATOR: The heat ON indicator lamp is lit when the
chamber heater is receiving power. The lamp will
normally flash when the chamber temperature is
at set point.
3. SET POINT ADJUSTMENTS: The temperature controller normally displays the
chamber temperature. To view or change the
temperature set point proceed as follows:
PRESS CONTROLLER
View set point
Decrease set point
Increase set point
A. Press and hold the star key and use either the up or down arrow
key to adjust the set point to the desired temperature. Release the
star key.
B. Allow at least 30 minutes for the chamber temperature to stabilize.
13

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific Lab-Line Imperial 3 and is the answer not in the manual?

Thermo Scientific Lab-Line Imperial 3 Specifications

General IconGeneral
BrandThermo Scientific
ModelLab-Line Imperial 3
CategoryAccessories
LanguageEnglish

Related product manuals