Thermo NESLAB HX 150 - Troubleshooting Guide; Troubleshooting Runaway Cold

Default Icon
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
START
Is
Setpoint
< actual
?
Calibration
error
NO
YES
Replace
valve
Valve is
mechanically OK
Test
STOP
Refer to
ELECTRICAL & CONTROL
section
Remove coil
from valve
Drawing 67.cfl
Rick Mills
13 March 1995
YES
YES
NO
Valve is
stuck open
Not a
Refrigeration
Problem
Problem is in
Controller or
sensor
See VALVE
REPLACEMENT
Does
unit
warm up?
YES
NO
UNSURE
YES NO
NO
Can you
get cool
lamp off? *
Is
cool lamp
on?
* by manipulating setpoint
Is
coil
on?
Normal
Response
Troubleshooting Chart: Runaway cold