Thermo NESLAB HX 150 - Troubleshooting Runaway Hot

Default Icon
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
START
Is
Setpoint
> actual
?
Normal
response
Calibration
Error
NO
YES
Replace
valve
Test
STOP
Refer to
ELECTRICAL & CONTROL
section
Drawing 68.cfl
Rick Mills
13 March 1995
YES
NO
Valve is
stuck closed
Not a
refrigeration
problem
See VALVE
REPLACEMENT
UNSURE
Can you
get cool
lamp on? *
Is
cool lamp
on?
YES
NO
NO
YES
NO
YES
NO
YES
NO
YES
*by manipulating setpoint
See TROUBLESHOOTING:
"WEAK COOLING"
Any
DC volts
on coil
?
Any
AC volts
on coil
?
Is
coil
on?
Problem is in
Controller or
Sensor
Is it
cooling
at all ?
Troubleshooting Chart: Runaway hot