6.3.5 Command FB
<id>FB<parameter><CR> Quick-start mode
Access: Read / write.
Description: The command FB is used to select quick-start mode.
Parameter description: 0: quick-start mode disabled
1: quick-start mode enabled
Value range: 0...1
Initial value: 1
6.3.6 Command HP
<id>HP<parameter><CR> Heating power Condensation protection
Access: Read / write.
Description: The heating capacity for the condensation protection of the
precipitation monitor is set by command “HP”. The indication is
stated in percent.
Parameter description: 5…8: Maximum responsivity of the precipitation monitor for
detection of the lowest precipitation intensities.
9…17: High responsivity, and at the same time reduction of the
responsivity for fog, mist and condensation.
18…22: Maximum insensibility against fog, mist and
condensation
Value range: 0...100
Initial value: 5