6 Programmers Reference
13
MTD415T Data Sheet Rev. 1.0
6.2.2 TEC Commands
Command
Explanation
Response Example
Programming
"Lx"
Sets the TEC current limit to x *)
Value range x: 200 to 2000 [mA]
Reading
"L?"
Reads the TEC current limit
[x<LF>][mA]
"A?"
Reads the actual TEC current
[x<LF>][mA]
x < 0: Heating; x > 0: Cooling
"U?"
Reads the actual TEC voltage
[x<LF>][mV]
*) This set value remains during the actual power-on cycle. Use the "M" command to save it to the non-volatile
memory.