18
MTD415T
MTD415T Data Sheet Rev. 1.0
6.3 Error Register and Safety Bitmask
The MTD415T has an internal 16 bit error register:
Bit Number
Event
0
Enable pin not set to L (GND)
1
Internal temperature too high
2
Thermal Latch-Up (TEC current at limit without temperature improvement)
3
Cycling time too small
4
No Sensor detected
5
No TEC detected (connection open)
6
TEC mispoled
7
(not used)
8
(not used)
9
(not used)
10
(not used)
11
(not used)
12
(not used)
13
Value out of range
14
Invalid command
15
(not used)
The error register can be read out using the "E?" command. The error register can be reset
using the "c" command or by setting the Enable pin to Off and On again.