MTD415T
MTD415T Data Sheet Rev. 1.0
6.2.3 Temperature
Command
Explanation
Response Example
Programming
"Tx"
Sets the set temperature to x *)
Value range x: 5000 to 45000 [10
-3
°C]
Reading
"T?"
Reads the set temperature
[x<LF>]
Value range x: 5000 to 45000 [10
-3
°C]
"Te?"
Reads the actual temperature
[x<LF>]
Command
Explanation
Response Example
Programming
"Wx"
Sets the set temperature window to x *)
Value range x: 1 to 32768 [mK]
Reading
"W?"
Reads the temperature window
[x<LF>][mK]
Command
Explanation
Response Example
Programming
"dx"
Sets the delay time between reaching the temperature window
and activating the Status output pin to x *)
Value range x: 1 to 32768 [sec]
Reading
"d?"
Reads the delay time between reaching the temperature
window and activating the Status output pin
[x<LF>][sec]
*) This set value remains during the actual power-on cycle. Use the "M" command to save it to the non-volatile
memory.