Calibration with a Weight:
In order to increase the accuracy of the machine you have to calibrate with a
known weight.
It is done as follows:
- remove load from the load cell (if there is any);
- set calibration factor to “0.
The display shows now:
(Do not be confused that the display shows the messages in German - this
program is used seldom, that is why we did not translate it into English. On the
right side of the page you see the messages translated into English).
Confirm with the [1] key, then you see:
Please care that there is no vibration - then press any key.
Now hang a known weight on the load cell, wait and then press any key.
Press any key.
The machine calculates now the factor automatically, it counts from 1 till 100%.
When the procedure is completed, the machine expects that the operator enters
the force that the load cell was loaded with.
Calibrate with weight
Weight calibre 1=ok
Remove load from the load
cell
Burden the load
The machine calculates the
factor
The machine calculates the
factor 15%
The pre-set force is 38.460
N
Enter the loaded force
mit Gewicht eichen
weight calibr. 1=ok
Lastzelle entlasten
weiter mit Taste
Last aufbringen
weiter mit Taste
Faktor bestimmen
15%
Last: 38.460 N
Neu:
Faktor bestimmen
weiter mit Taste