31
Table of orders
Order
Function
Example/
Syntax
Result
End of command is $ or #10
*
Decimal separator is always '. ' (point)
'A'
Actual readings
A$
Force ; Way; Way (from
extensometer or LVDT or MFE900
only when is enabled)
'B'
Last readings ( rupt. value)
B$
F rupt X rupt
'D xxx'
Set F delta
D20$
F delta=20N
'DA xxx'
Ask for “Auto delta + 20%”
DA$
Return Auto delta=21N
'E'
Return to zero position (if Ref mode
>0 Return to “Start pos”.
E$
Return button (to 0,00mm)
'F xxx'
Set limit of force in test (F limit S)
F200$
F limit S=200N
'H'
Stop machine (or break CFR)
H$
Stop button
'I?'
Ask for Way display mode (if
Extensometer or LVDT or MFE900 is
enabled)
I? $
“I:0” – standard Way display
“I:1” – Way from Extensometer on
LCD
“I:3” – Way from LVDT on LCD
“I:4” – Way from MFE900 on LCD
'Ix y'
Set Way display mode and L0 value
(if Extensometer or LVDT or MFE900
is enabled).
I1 25.00$
Way from Extensometer on LCD
and L0=25.00 mm
'J xxx'
Set return speed
J200$
VR=200 mm/min
'K0'
Set normal mode
K0$
Machine in Normal mode
'K1'
Set Constant Force Regulation mode
K1$
Machine in CFR mode
'K?'
Ask for status Constant Force
Regulation mode (active or not)
K? $
“K1$” - Machine in CFR mode
“K0$” - Machine not in CFR mode
'L xxx'
Set end position for test
L100$
X=100mm
'M'
Last readings (max. value)
M$
F max X max
'N'
Read nominal load
N$
2000 (Load cell = 2000N)
'O'
Reset Load cell and position counter
O$
0,0N 0,00mm
'O1'
Reset position counter only
O1$
2,0N 0,00mm
'O2'
Reset Force only
O2$
0,0N 7,34mm
'O3'
Reset Time Stamp (option)
O3$
'O4'
Reset LVDT (when LVDT enabled)
O4$
'O5'
Reset MFE900 counters (when
MFE900 enabled)
O5$
'OR'
Reset Reference offset (when
Ref mode >0)
OR$
0,0N 7,34mm
'P xxx'
Set pre-force (F0)
P0.4$
*
F0=0.4N
'PP xxx'
Set pre-force for perforce speed (in not
the same as ‘F0’.
PP0.4$
*
Pre-force=0.4N
'Q xxx'
Set constant force to keep (in CFM)
Q200$
Force to keep = 200N
‘QH xx’
Set the stiffness factor for CFR
Where 0 <= xx <= 10
QH5$
Stiffness=5
(min 0, max 10)
‘QT xxx’
Time of Constant Force Regulation (1-
604800 sec)
QT3600$
After 1 hour End of CFR
‘R’
Begin of test (start the test)
R$
As ‘START’ button
‘RS’
Begin of silent test (start the test)
RS$
This test sent no data, host can ask