Setting T[p1] = F[p2] is forbidden!
A gap between the thresholds being too small will result in
flickering of the relay and can even cause damage to the relay!
Setting E: Relay closes in case of a device error.
Setting !E: Relay opens in case of a device error.
Setting U: Relay closes in case of pressure underrange.
Setting !U: Relay opens in case of pressure underrange.
Setting O: Relay closes in case of pressure overrange.
Setting !O: Relay opens in case of pressure overrange.
Setting C: Relay closes when cold cathode is switched on.
Setting !C: Relay opens when cold cathode is switched on.
Setting T0: Open relay by software command (off).
Setting T1: Close relay by software command (on).
The control LEDs on top of the transducer will be on as soon
as the related relay is activated.