EasyManua.ls Logo

Tianlong 96E - Run Monitoring

Tianlong 96E
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
Reminding: the temperature center setting range is 35.5°C ~99.5°C.
Reminding: the temperature offset setting range is 0.5°C ~20.0°C.
Reminding: Gradient settings only apply to Genesy 96T.
3.2.3 Run Monitoring
After finishing the experiment setting, users could press the < Run Experiment > function
key to start running the current experiment, the system will enter the run monitoring interface,
as shown in figure 3-4.
Figure 3-4 Run Monitoring Interface
The run monitoring interface is consisted of two parts: run status bar and run monitoring
diagram. Users could monitor the real-time status of the current running experiment on this
interface.
1) Run Status Bar: display the real-time status of the current running experiment, including
the Remaining Time, the current Stage, Cycle and Step, as shown in figure 3-4-1.
Figure 3-4-1 Run Monitoring Interface - Run Status Bar