VI
Biohazard: user should consider the abandoned Genesy thermal cycler as biological
contaminated material and comply with the local or national applicable regulations to
complete the instrument disposal. Before disposal or recycle the instrument, please
completely clean and disinfect the instrument.
Biohazard: user should only use reagents and consumables that are within their
expiration date.
7. Instrument Operation Safety
Caution: no person except the XATL Co., Ltd. professional maintenance personnel is
allowed to open the instrument, replace any part or disassemble machine. Do not replace
any part while the instrument is connected to the power supply.
Caution: do not drop or damage the instrument, please handle it carefully.
Reminding: please cut off the power and restart the system if the instrument is not
working properly.
Reminding: when only needs one or two single centrifugal tube(s) for experiment, please
place one empty centrifugal tube in each corner of the block in order to protect the
centrifugal tube(s) from damage because of over-pressure.
High-Temp: the temperature of hot lid may raise up to 105°C during the process of
experiment, please pay attention to avoid high-temperature hazards.
Prohibit: never open the hot lid while the instrument is still running, which may affect the
experiment result.
8. Other Details
8.1 Grounding
For protection against electric shock hazards, the Genesy thermal cycler must be grounded
properly. The power cord provided is with a standard three pin plug, please plug it into an
appropriate three-wire grounded receptacle for operation safeties.
8.2 Power Supply
Before connecting the AC power supply, please ensure the consistent of the Genesy thermal
cycler required voltage and the power supply voltage, and make sure that the rated load of
receptacle is no less than the requirement of the instrument.