7
5.3 PRESS <TARE> TO CONFIRM
5.4 NOW DISPLAY INTERNAL NUMBER 0IF IT IS NOT 0, PRESS <ON/ZERO> TO ZERO
5.5 PLACE MORE THAN 10% MAX.CAPACITY STANDARD WEIGHTS ON PLATE, AFTER INTERNAL NUMBER
IS STABLE, PRESS <TARE>
TO INPUT THE VALUE OF WEIGHTSPRESS <SET/0~9> TO ADD 1PRESS <TARE> TO INPUT NEXT
VALUE, TILL TO FINISH, IT
SHOWS CAEnd ONE SECOND, RETURN TO WEIGHING. PLEASE ATTENTION THE PLACE OF DECIMAL
WHEN INPUTING
5.6 IF DISPLAY Err03MEANS CALIBRATION FAULT. THE REASON IS: WEIGHTS ARE TOO SMALL OR
INPUT WRONG VALUE.